The following requirements apply to tests performed in laboratories under the environmental reference conditions outlined in 8.1.
The EUT’s shall be placed on a ground reference plane and shall be insulated from it by an insulating support 0.1 m ± 0.01 m thick.
In the case of table-top equipment, the EUT should be located 0.8 m ± 0.08 m above the ground plane (see figure 7).
The reference ground plane shall be a metallic sheet (copper or aluminum) of 0.25 mm minimum thickness; other metallic materials may be used but they shall have 0.65 mm minimum thickness.
The minimum size of the ground plane is 1 m x 1 m. The actual size depends on the dimensions of the EUT.
The reference ground plane shall project beyond the EUT by at least 0.1 m on all sides.
The reference ground plane shall be connected to the protective earth (“ground” in US terminology).
The EUT shall be arranged and connected to satisfy its functional requirements, according to the equipment installation specifications.
The minimum distance between the EUT and all other conductive structures (e.g. the walls of a shielded room), except the ground plane beneath the EUT, shall be more than 0.5 m.
The EUT shall be connected to the earthing system in accordance with the manufacturer’s installation specifications; no additional earthing connections are allowed.
The connection of the test equipment ground cables to the ground reference plane and all bondings shall provide minimum inductance.
Coupling devices shall be used for the application of the test voltages. They shall be coupled to the lines between the EUT and the decoupling network or between two units of equipment involved in the test.
Using the coupling clamp, the minimum distance between the coupling plates and all other conductive structures, except the ground plane beneath the coupling clamp and beneath the EUT, shall be 0.5 m.
The length of the signal and power lines between the coupling device and the EUT shall be 1 m or less.
If the manufacturer provides a non-detachable supply cable more than 1m long with the equipment, the excess length of this cable shall be gathered into a flat coil with a 0.4 m diameter and situated at a distance of 0.1 m above the ground reference plane. The distance of 1 m or less between EUT and the coupling device shall be maintained.
Examples of the test set-up for laboratory tests are given in figure 7.